Our Services

Radiotherapy Calibrations

x-ray photograph

ARPANSA provides calibration services for instruments used to measure radiation in various commercial, medical and public sector applications.

This Page:

Primary standards

ARPANSA maintains primary standards for ionizing radiation (Table 1) and perform calibrations of therapy dosimetry equipment that are traceable to these standards. Calibrations at 60Co are used as the basis for megavoltage dosimetry, and are suitable for use with protocols TRS-277, TRS-381, TRS-398 and the “ACPSEM Protocol (2nd Edition)”.

Contact our Ionizing Radiation Standards Section for assistance.

Table 1: Australian primary standards maintained at ARPANSA
Quantity Beam quality Primary standard
Air kerma / exposure 60Co Graphite cavity chamber
Absorbed dose to graphite (1) 60Co Graphite calorimeter
Air kerma / exposure Medium-energy x-rays (50-300 kVp) Medium-energy free air chamber (MEFAC)
Air kerma / exposure Low-energy x-rays (20-50 kVp) Low-energy free air chamber (currently operating as a secondary standard traceable to the MEFAC)

Note: absorbed dose to water is calculated from the absorbed dose to graphite using the photon fluence scaling theorem.

Air kerma calibration in 60Co

  • Ionization chamber is calibrated with the build-up cap on
  • The reference point is the geometrical centre of the chamber
  • The reference point is placed at 99.3 cm from the source where the air kerma rate is known
  • The beam is square with approximate size 10 cm x 10 cm
  • The air kerma rates varies as the 60Co decays. Over the 10 year life of the source the rate varies from about 7 mGy/s down to about 1 mGy/s.
  • The calibration coefficient is NK

Absorbed dose to water calibration in 60Co

  • Ionization chamber is calibrated with the build-up cap removed and (if the chamber is not waterproof) placed in a 0.5 mm thick Perspex sleeve in a water phantom. The reference point is the geometrical centre of the chamber.
  • The reference point is placed 5 cm beyond the surface of the water phantom where the absorbed dose to water rate is known
  • The source-surface distance (SDD) is 100 cm
  • The beam is square with approximate size 10 cm x 10 cm
  • The phantom is a custom built Perspex tank of dimensions 40 cm x 40 cm x 40 cm. The beam enters the side through a 2 mm thick Perspex window. The absorbed dose at 5 cm depth is corrected for the non-water equivalence of this window.
  • The absorbed dose to water rates varies as the 60Co decays. Over the 10 year life of the source the rate varies from about 7 mGy/s down to about 1 mGy/s.
  • The calibration coefficient is ND,w

Polarity and recombination corrections

  • Polarity correction measured at 60Co by reversing the polarizing voltage of the ionization chamber.
  • Recombination correction at 60Co is checked using the two-voltage method (for therapy ionization chambers, this correction is usually less than 0.1% at 60Co).

Medium-energy kilovoltage x-rays

  • The ionization chamber is calibrated in a horizontal beam
  • The source-detector distance is 1331 mm and the beam diameter approximately 11 cm
  • The build-up cap is removed (except in certain cases for ST17)
Table 2: Medium energy therapy radiation qualities available from the Seifert x-ray generator for a source-detector distance of 1331 mm, nominal beam diameter 11 cm
Beam ID kVp Tube current (mA) Additional filtration Air kerma rate (mGy/s) First HVL (mm Al) First HVL (mm Cu) Average energy (keV)
Al(mm) Cu (mm) Sn(mm)
ST1 30 10 0.32     0.4 0.35   19
ST2 50 10 0.32     0.8 0.56   26
ST3 75 20 0.32     2.5 0.77   34
ST4 100 20 0.32     3.1 1.01   42
ST5 75 20 1.05     1.4 1.50   37
ST6 100 20 1.05     2.0 2.06   45
ST7 135 20 1.05     2.7 2.93   54
ST8 100 20 3.45     1.0 3.94 0.14 50
ST9 150 20 3.45     1.8 5.81 0.27 62
ST10 135 20 1.05 0.25   1.0 8.42 0.47 66
ST11 135 20 2.02 0.27   0.8 8.86 0.51 68
ST12 150 20 2.02 0.27   1.0   0.60 71
ST13 180 20 2.02 0.54   1.1   0.99 84
ST14 250 15 2.02 0.27   2.2   1.17 89
ST15 250 15 2.02 0.54   1.7   1.53 97
ST16 250 15 0.99 1.86   0.9   2.52 123
ST17 300 12 1.95 3.94 0.25 0.6   3.98 165

Low-energy kilovoltage x-rays

  • The ionization chamber is calibrated in a horizontal beam
  • The source-detector distance is 782 mm and the beam diameter approximately 11 cm
  • The build-up cap is removed (if relevant)
  • Suitable for parallel plate chambers
Table 3: Low/medium energy therapy radiation qualities available from the RT100 x-ray generator for a source-detector distance of 782 mm, nominal beam diameter 10 cm
Beam ID Tube kVp (kV) Tube current (mA) Additional filtration Air kerma rate (mGy/s) First HVL (mm Al) Average energy (keV)
(mm Al) (mm Cu)
RT1 20 10 0.15   2.4 0.125 12
RT2 30 10 0.30   2.3 0.23 18
RT3 37 10 0.40   2.5 0.38 21
RT4 45 10 0.55   2.5 0.58 24
RT5 55 10 0.78   2.6 0.87 29
RT6 70 10 1.25   2.6 1.40 35
RT7 100 8 1.70   2.9 2.34 44
RT8 100 8 1.02 0.25 0.7 6.67 57

Air kerma calibrations for 137Cs and 192Ir

ARPANSA does not maintain primary standards at these qualities. Instead, ionization chambers can be calibrated at 60Co and at the highest energy X-ray beam (ST17) with the build-up cap on. The calibration coefficient at the energy of interest can then be found by interpolating the results at 1,250 keV (60Co) and 164 keV (ST17).

Note: ANSTO provide a service for the calibration of well-chambers for selected radionuclides.

Electrometer calibrations

  • Charge sensitivity, traceable to Australian electrical standards
  • Linearity
  • Ratio of ranges
  • Ratio of integral and differential measurement modes

Additional notes about ARPANSA therapy calibrations

  1. During the calibration of a therapy dosemeter, the ionization chamber is calibrated separately. This allows the ionization chamber to be used with a different electrometer, and enables drifts etc related to the chamber to be separated from electrometer effects. During at least on of the 60Co measurements a combined calibration is performed as a check.
  2. A 90Sr check source measurement with the client’s source if supplied. ARPANSA has its own 90Sr check sources which may also be used.
  3. Electrometer calibrations are made using the client’s ionization chamber. The current from the ionization chamber in a 60Co beam is measured using a custom-built current integrator, which is traceable to national standards of electrical measurement. The electrometer is calibrated using the known current delivered over a time interval either with the ARPANSA source timer, or using the internal timer of the electrometer.

A series of fact sheets concerning radiotherapy calibrations made at ARPANSA can be found on the equipment calibration page.

Top of Page